Specifications
AM25T Solid State Multiplexer 
2 
OPERATING TEMPERATURE 
Standard: -40°C to +85°C 
 RTD accuracy ±0.4°C 
OPERATING HUMIDITY 
Noncondensing: 0 − 95% 
DIMENSIONS 
Length:  30.5 cm (12.0”) 
Width:  5.10 cm (2.0") 
Depth:  15.3 cm (6.0”) 
WEIGHT 
0.91 kg (2.0 lbs) 
2.7 kg (6.0 lbs); Shipping 
EXPANDABILITY (nominal)** 
4 AM25Ts  per CR1000 
6 AM25Ts  per CR23X 
4 AM25Ts  per CR10(X) 
6 AM25Ts  per 21X 
7 AM25Ts  per CR7 725 Card 
MAXIMUM CABLE LENGTH 
Multiplexers can be located up to 305 m (1000 ft) from the datalogger. When 
lighting protection is required, do not exceed 152 m (500 ft) in cable length. The 
spark gaps will not fire if the heavy ground wire is longer than 152 m (500 ft). 
TYPICAL RELAY RESISTANCE 
500 Ω 
MAXIMUM SWITCHING CURRENT 
25 mA; Switching currents greater than 25 mA will damage the relays and render 
them unusable. 
**Assumes sequential activation of multiplexers and that each datalogger channel 
is uniquely dedicated. If your application requires additional multiplexing 
capability, please consult CSI for application assistance. 
1.2 Physical Description 
The AM25T is housed in an anodized aluminum case with a cover that will 
help reduce temperature gradients across the AM25T's terminal strips 
(Figure 1). 