Instruction manual
CNR4 Net Radiometer 
The CNR4 outputs four voltages that typically range from 0 to 15 mV for the 
pyranometers, and ± 5 mV for the pyrgeometers. A differential voltage 
measurement is recommended because it has better noise rejection than a 
single-ended measurement. If differential channels are not available, single-
ended measurements can be used. The acceptability of a single-ended 
measurement can be determined by simply comparing the results of single-
ended and differential measurements made under the same conditions. 
Additionally, one voltage excitation channel and one single-ended analog 
channel are required to make the temperature measurement of the sensor body, 
using the thermistor. 
7.3.1  Sensor Sensitivity 
The CNR4 comes with four different sensor sensitivity values for four separate 
probes. The CNR4 sensor comes with two copies of its ‘Certificate of 
Calibration’ by the manufacturer. They show the sensor serial number and 
sensitivity values for four individual probes: one copy for pyranometers, and 
another copy for pyrgeometers. The serial number and sensitivity values are 
also shown on a label affixed to the bottom of the sensor. If you choose to 
attach the CNF4 heater/ventilator unit to the CNR4, the label showing the 
serial number and sensitivity values will be covered. After attaching the CNF4 
heater/ventilator, affix the extra label to the bottom of the CNF4 in a visible 
location. The extra label containing the serial number and sensitivity values is 
supplied with the purchase of the CNR4. Please refer to Appendix D, CNF4 
Heater/Ventilator, for more details. 
The sensor sensitivity is in µV/(W/m
2
). This needs to be converted into 
(W/m
2
)/mV to be used as a multiplier parameter inside the datalogger program. 
To convert the units, divide the sensor sensitivity value into 1000. For 
example, if the sensitivity is 7.30 µV/(W/m
2
), the multiplier is 1000/7.3 = 
136.99 (W/m
2
)/mV. 
8.  Troubleshooting 
If there is no indication as to what may be the problem, start performing the 
following “upside-down test”, which is a rough test for a first diagnosis. It can 
be performed both outdoors and indoors. Indoors, a lamp can be used as a 
source for both short-wave and long-wave radiation. Outdoors, one should 
preferably work with a solar elevation of more than 45 degrees (45 degrees 
above horizon) and under stable conditions (no large changes in solar 
irradiance, and preferably no clouds). 
1.  Measure the radiation outputs in the normal position. Record the 
measured values when the signals have stabilized, i.e. after about three 
minutes. 
2.  Rotate the instrument 180 degrees, so that the upper and the lower sensors 
are now in the reverse orientation as to the previous position. 
3.  Measure the radiation outputs once more. Record the measured values 
when the radiometers have stabilized. 
4.  The computed net radiation values in rotated position should be equal in 
magnitude but only differing in sign. In a rough test like this, deviations 
of ± 10 % can be tolerated. If deviations greater than this are encountered, 
additional testing is warranted. 
22 










