Instruction manual

CS451/CS456 Submersible Pressure Transducer
when digital communications break down, resulting in either no response or
excessive SDI-12 retries and incorrect data due to noise problems. (Using
SDI-12 commands like aMC!, which adds a CRC check, can significantly
improve incorrect data issues.)
8.3 Measuring Multiple SDI-12 Sensors
Up to ten CS451/CS456 sensors or other SDI-12 sensors can be connected to a
single datalogger control port. Each SDI-12 device must have a unique SDI-12
address of 0 and 9, A to Z, or a to z. See Appendix D, SDI-12 Sensor Support
(p. D-1), for more information.
9. Maintenance
Campbell Scientific recommends that the CS451/CS456 be factory recalibrated
and checked every 24 months. Before a CS451/CS456 sensor is sent to
Campbell Scientific, the customer must get an RMA (returned material
authorization) number, and fill out the Declaration of Hazardous Material and
Decontamination form.
The CS451/CS456 has no user-serviceable parts. Cable can be damaged by
abrasion, rodents, sharp objects, twisting, crimping or crushing, and pulling.
Take care during installation and use to avoid cable damage. If a section of
cable is damaged, it is recommended that you send your sensor back to replace
the bale harness assembly.
Periodic evaluation of the desiccant is vital for keeping the vent tube dry. The
CS451/CS456 ships with the desiccant tube attached. To assess the
effectiveness of the desiccant, use one of the following:
x The desiccant in the tube changes color from blue to pink when the
drying power is lost.
x The Enclosure Accessory Humidity Indicator Card (pn 28878).
9.1 Every Visit
x Collect data.
x Visually inspect wiring and physical conditions.
x Check indicating desiccant or enclosure humidity indicator; service if
necessary.
x Check battery condition (inspect physical appearance and use a keyboard
display, PDA, or laptop to view the battery voltage).
x Check all sensor readings; adjust transducer offsets if necessary.
x Check recent data.
9.2 Every Two to Three Years or on a Rotating Schedule
x Send the CS451/CS456 in for inspection.
17