Use Instructions

CanarE System
Page | 19 Instructions for Use
Rev. 4
Figure 10: Attaching CanarE with Strap
b. If using the self-adhesive tape:
i. Wrap the self-adhesive tape around the CanarE and patient’s knee
multiple times with overlap to secure the CanarE.
NOTE: When pulling the self-adhesive tape tight to secure the CanarE to the
patient, the CanarE may slightly shift on the patient’s knee. Make sure the
CanarE placement on the patient’s knee has not changed. If the CanarE has
shifted, readjust its position to the correct location. If the position is not
corrected, this may affect the CanarE data collection.
Instruct the patient to walk with the CanarE attached to their knee. Observe and
make sure the CanarE is secure and is not shifting position while the patient walks.
Adjust the CanarE as needed.
NOTE: Ask the patient if the CanarE feels uncomfortable or too loose and make
adjustments.
8.3. Performing Data Collections with the CanarE
Enter the CanarE serial number, found on the patient’s device label, into the Clinic
App (see the
Clinic Base Station System IFU).
Follow the Clinic Base Station System IFU to perform data collection using the
CanarE.
During data collection, the CanarE LED will indicate when the CanarE is active. The
LED status is defined in
Table 5.
NOTE: See troubleshooting steps in Section 14 if the CanarE LED is a constant
yellow.