Use Instructions

Page |
34
2. Opens the package, inspects the equipment, and identifies the enclosed Quick Start
Guide and Patient Manual.
3. Turns on the patient’s personal computer with USB connection, opens a web browser,
navigates to the Canary website, and logs into the patient’s account, which was
created previously.
4. Follows the instructions in the Quick Start Guide and Patient Manual to set up the
Base Station.
11.4. Day of Surgery
The following provides instructions on how to set up, test, and activate the CTE implant. It also
tells you how to link the CTE implant and associated Zimmer knee components to the specific
patient.
This information is a summary. For a complete description of this process, including example
screenshots, refer to the Canary Medical Surgical Technique, document # K01-CTE-300005,
provided at www.canarymedical.com.
11.4.1. Preoperative: CTE Implant Self-Test and Sensor Check
NOTE: The OR App will not function and you will not be able to log in without first connecting
the OR Base Station to the Laptop.
1. Turn on the Canary Laptop and ensure it has a Wi-Fi connection
2. Gather a CTE implant, the OR Base Station Unit, Bar Code Scanner, and USB data and power
cables.
a. Check the expiration date on the CTE package to ensure the implant is not expired.
WARNING: To avoid potential patient injury, do not use the CTE implant if it is expired.
NOTE: Do not open the CTE package at this time.
3. Set up the OR Base Station system in or near the OR but outside the sterile field, using the
steps below:
a. Place the Base Station stand on a flat surface. Place the Base Station into the stand
as shown in Figure 16. Insert the provided screw through the groove in the stand into
the Base Station. Tighten the screw with a screw driver.
Physician Instructions for Use
DRAFT