Use Instructions

Page |
36
f. Ensure the OR Base Station and Laptop are connected. This is indicated by the green
“Base Station” icon at the top right of the Laptop screen.
g. Ensure an Internet connection has been established and the OR App is connected to
the Internet. This is indicated by the green “Internet” icon at the top right of the
screen. If the OR App is not connected to the Internet, check the time of last sync. If
it has been some time since this has been done, move to a location and attempt to
connect to the Internet.
h. Click on the Canary logo in the center of the screen.
i. Enter your authorized Username and Password and click the “Login” button.
j. On the Main Menu screen, under “Preoperative,” click on “Setup Implant.”
k. Enter or scan the CTE serial number from the implant package and click “OK.”
l. The Base Station initiates communication with the CTE implant. Click “OK” when this
process is finished.
m. Click on Self-Test.
n. Click on Sensor Check.
o. When the Self-Test and Sensor Check are successful (shown by the “implant status”
indicators on the screen), the CTE implant is ready to be used for the patient’s TKA
surgery. If the Self-Test and Sensor Check have been run successfully within the last
24 hours, the status will reflect that, and the user can elect to not repeat these tests.
11.4.2. Intraoperative: Tibia Resection, Preparation, and Trialing
The tibia resection and preparation is done similarly to the Zimmer Biomet Persona® technique.
However, the CTE with CHIRP System surgical technique deviates from the Persona Knee
Surgical Technique in order to properly place the CTE implant.
Refer to both the Canary Medical CTE with CHIRP System and Zimmer Biomet Persona Knee IFUs
and Surgical Techniques for the detailed technique steps. See Section # 2 of this IFU for
document references.
The steps to complete the tibia resection and preparation are summarized below.