Use Instructions

Page |
37
CAUTION: Perform the tibia resection using the Canary Medical Tibia Resection Cutting Guide.
The Zimmer Biomet Persona Tibia Resection Cutting Guides have not been tested for use with
the Canary Tibial Extension (CTE).
1. Use the Persona instrumentation to properly align and size the tibia plate
2. Create a pilot hole using the Zimmer Biomet Drill Guide and Persona Tibial Drill Bit. Then
use the Canary Medical Drill Bit and Drill Guide to create a cavity in the IM canal for the
CTE implant and cement mantle.
3. Broach the tibia using the Persona instrumentation to create a cavity for the Persona Tibia
Plate.
4. Assemble the CTE Provisional to the Persona Tibia Plate Provisional and insert the
provisional construct into the prepared tibia to assess the fit of the construct. Confirm
the final implant sizing.
11.4.3. Intraoperative: Activate the CTE implant
After the CTE implant has been introduced into the sterile field, and before assembly with the
Persona tibial plate, activate the CTE implant using the OR PC Application and OR Base Station.
1. Go to the Main Menu screen.
2. Under “Intraoperative,” click on “Activate Implant.”
3. Click in the text box and enter or scan the CTE Serial Number. Click “OK.”
4. Click “Activate.”
5. The Activation Status on the screen will change from “Not Activated” to “Activated.”
11.4.4. Intraoperative: Connection of Canary CTE to Zimmer Biomet Tibial Baseplate and
Implantation
NOTE: In order to attach the Canary CTE to the Zimmer Biomet Persona® Knee tibial plate, the
following Canary Medical instrumentation is required:
Physician Instructions for Use
DRAFT