16.2. Base Station Units and Accessories
Maintain the Base Station Units and Accessories by keeping them clean, dry, and free of
surrounding clutter.
Home Base Station: Dust with a clean, dry, soft cloth.
OR Base Station: The following solutions have been tested for safe cleaning of the OR Base
Station unit, part # 43-5570-002-14 only. Wipe the OR Base Station Unit with a soft cloth
dampened with one of these solutions.
1. 1:10 Bleach Solution
2. 1:10 Mild Detergent Solution
3. Ammonia Solution (RTU)
4. 70% Isopropyl Alcohol
5. 0.5% Hydrogen Peroxide
6. Phenolic germicidal detergent solution (RTU)
7. Iodophor germicidal detergent solution (RTU)
WARNING: Do not immerse the base stations in water or cleaning agents as this could cause
electrocution.
16.3. SOFTWARE MAINTENANCE
All software updates will be initiated by Canary Medical.
17. SERVICE LIFE
17.1. CTE
The service life of the CTE is estimated to be at least ten years based on; a) accelerated discharge
testing of the batteries, b) bench testing of current consumption by the CTE electronics, and c)
assuming no more than 1000 sampling days are programmed into the CTE. However, the CTE is
designed for up to 20 years of service life.
17.2. Base Station Units and Accessories
The Base Station Units and Accessories are intended to last for 3 - 5 years and can be replaced in
case of them being non-operational.