Internal Battery:
Type Lithium-Ion
Operating time 70 hours minimum
Charging time 13 to 17 hours
Lifetime 300 cycles
CAUTION: Risk of explosion if battery is incorrectly replaced. Replace only with the same or
equivalent type recommended by the manufacturer. Dispose of used batteries according to the
manufacturer's instructions
Communication:
Frequency 915 MHz (US)
Range 1000 m (unobstructed)
300 m (inside building)
Dimensions
Length 160 mm
Width 160 mm
Height 80 mm
Material Aluminium alloy
Color beige
Cleaning Resistance Alcohol (90% solution)
Certifications & Approvals:
CEM/Radio 915MHz FCC 15-247 (US) FCC ID: VPM-RS-915
5.3. NORMAL POWER ON/OFF SEQUENCE
The following procedure outlines the normal operating sequence of the Supervisor/Repeater module.
• For a Supervisor-PC or Repeater-PC, connect the Supervisor/Repeater module to the host
computer USB port using the supplied data communication cable. Note that a Secondary-
Repeater or Supervisor is operated as a standalone module, and would not be connected to a
PC.
• Switch on the Supervisor/Repeater module by pressing and briefly holding the power button until
all four status LEDs turn on.
• If external power is connected to the module, the Blue status LED will turn on briefly and the
Supervisor/Repeater module will indicate its charging status by turning on the appropriate color
LED as follows:
Green = 100% capacity (charge complete)
Orange = less than 99% capacity
Red = less than 10% capacity