LOW BATTERY INDICATOR
The presence of the “LOW BATTERY” indicator (not
flashing) on the display indicates that the battery is
nearly low.
If the indicator remains displayed for several consecutive
infusions, the “LOW BATTERY” message is displayed,
accompanied by a beep repeated approximately every
10 seconds.
In these circumstances the pump can no longer be used
and the battery must be replaced.
SECTION 5
WARNINGS
Do not use rechargeable batteries.
Using other types of battery than lithium CR 123 A batteries could cause the
pump to malfunction.
The battery life can be influenced by the age and the temperature and
circumstances of its use and storage.
Ensure you always have a replacement battery available for use.
If the pump is left inactive for long periods (1-2 months), you are advised to
remove the battery.
After you have inserted the battery, the pump runs a self-diagnosis test
during which it will emit brief audio signals and display all of the icons and
indicators.
When you have finished changing the battery, check that the compartment
is properly closed.
NOTES
During battery replacement, when in the OFF or StoP states, the pump retains
the current settings and the position of the pusher in its memory.
If the battery needs to be changed during an infusion the pump must be in the
StoP state.
If the battery is removed with the pump in the ON state, the pump is
automatically re-initialized, i.e. the pusher is withdrawn until it reaches the
“machine zero” position (pusher in contact with the syringe support) and, then
repositioned to start an infusion, displaying OFF on the display.