DISPLAY
ACOUSTIC
SIGNAL
ERROR
DESCRIPTION
CORRECTIVE
ACTION
Intermittent acoustic
signal repeated
approximately every
10 seconds.
When a battery is inserted
and at the start of every
infusion, the pump
performs a check of the
settings in the memory. If
an error is found, the value
in error is replaced by the
default value, the pump
motor is locked and the
error is indicated both on
the display and audibly.
Initialize the
pump.
Intermittent acoustic
signal repeated
approximately every
10 seconds.
Anomaly in the safety
circuit which drives the
pump motor. If an error is
found, the pump locks and
the error is indicated.
Initialize the
pump.
Intermittent acoustic
signal repeated
approximately every
10 seconds.
Anomaly in the pusher
mechanism.
Initialize the
pump.
Intermittent acoustic
signal repeated
approximately every
10 seconds.
Mechanism blocked
because of an occlusion in
the infusion line.
Eliminate the
cause and press
the
button.
See page 28.