The displayed error messages (from Er,2 to Er,11 and OCCL) are
accompanied by an acoustic signal and the system stops.
To initialize the device, remove the battery when the pump is in an error
state or the ON state and reinsert it after 10/15 sec. If the error is detected
again after the corrective action or initialisation of the device, contact the
CANÈ S.p.A. Technical Support Service.
Following the display of error message Er,8 and the successive initialisation,
the system reverts to the factory settings (see page 29): in this event the
pump settings prescribed by the doctor should be re-entered.
Error messages Er,2 and Er,7 are accompanied by the flashing red LED.
WARNINGS