User guide
27
MAN 01/EN/01 CRONO S-PID 50 02/13
• The displayed error messages (from Er,2 to Er,11 and OCCL) are
accompanied by an acoustic signal and the system stops.
• To initialize the device, remove the battery when the pump is in an error
state or the ON state and reinsert it after 10/15 sec. If the error is detected
again after the corrective action or initialisation of the device, contact the
CANÈ S.p.A. Technical Support Service.
• Following the display of error message Er,8 and the successive initialisation,
the system reverts to the factory settings (see page 29): in this event the
pump settings prescribed by the doctor should be re-entered.
• Error messages Er,2 and Er,7 are accompanied by the flashing red LED.
WARNINGS
SECTION 7