User guide

28
MAN 01/EN/01 CRONO S-PID 50 02/13
POST-OCCLUSION BOLUS
The occlusion alarm is given when the pump detects excessive back pressure
in the infusion line. This back pressure must be removed without accidentally
releasing a post-occlusion bolus, which could cause serious harm to the patient.
The volume of a CRONO S-PID 50 post-occlusion bolus, considering only the
combined volume of the pump and the-reservoir is about 1.8 ml.
The volume of the bolus released after an occlusion can vary, depending
on the type of catheter, the infusion set and all the other components that
comprise the infusion line.
Another element that could affect the volume of the released bolus after an
occlusion is the presence of any air in the system.
After the occlusion alarm is given, disconnect the infusion set from the patient
to avoid a post-occlusion bolus being administered to the patient.
WARNINGS
SECTION 7
INFUSION SET OCCLUSION
The pump is designed to recognize when the administration
of a pharmaceutical has been interrupted by external means,
such as, for example, the kinking of the infusion set tube
and consequent occlusion. In these circumstances, the
pump stops the infusion: the display indicates that there is
an occlusion, accompanied by a brief acoustic signal. The
pump then continues beeping every 10 seconds. While the
system is still occluded, the drug is not administered: to
recommence the infusion, press the
button, after having
removed the cause of the occlusion.
The cause of the occlusion is to be found along the infusion line and at the
point of injection.
To avoid or reduce the incidence of occlusions, you are advised to use an
infusion set with anti-kinking tubes.
NOTES