User guide

33
MAN 01/EN/01 CRONO S-PID 50 02/13
PUMP INITIALISATION
If the battery is removed when the display shows OFF when
you reinsert the battery the pump runs the initialisation
sequence, during which it:
1. Runs a self-diagnosis test, emitting a series of brief
acoustic signals, flashing the red LED and displaying all
the indicators and icons on the screen.
2. The display shows the previously selected
programming mode.
3. The display shows OFF.
The pump is supplied with a new battery already inside the pump.
For instructions on how to install the battery, see page 22.
You are recommended to initialize the pump if it is left unused for a long
period (more than 1 - 2 months) and the battery is not removed.
If the battery is removed when the display shows StoP, when you reinsert
the battery the pump runs a self-diagnosis test (as shown in Step 1) and
then the display will show StoP.
SECTION 10
NOTES
The setting of the pump is the responsibility of the doctor, who will choose the
parameter values best suited to the therapy required for the patient.
WARNING