User guide
34
MAN 01/EN/01 CRONO S-PID 50 02/13
If the battery is removed when the display shows OFF when
you reinsert the battery you can set the programming mode
of the pump:
1 - by flow rate, expressed in ml/h if you select “F” or,
2 - by time, expressed in hours and minutes if you select “t”.
Procedure:
1 - Remove the battery when it is in the OFF state and
reinsert it.
2 - The display shows all the symbols.
3 - At the same time, the device carries out an
auto-diagnosis during which it emit acoustic signals.
4 - The display will then show the programming mode “F”
(flow rate) or “t” (time): pressing the
button makes
the symbol flash for 9 seconds; you can move from one
option to the other using the
button.
5 - Not pressing these buttons for 9 seconds will mean
that the device will memorise the mode that has been
selected.
6 - Then, the display shows OFF.
PUMP SETTINGS WHEN INSERTING BATTERY
NOTE
Setting the programming mode is only possible with the L 0 access level and
only at the start of the infusion.
Choosing whether to program the pump by flow rate or by time is the
responsibility of the doctor, who will decide on the most suitable method.
WARNING
SECTION 10