User guide

36
MAN 01/EN/01 CRONO S-PID 50 02/13
SECTION 10
When the settings lock is on (L 1), if any attempts are made to change the
parameter then the display will show the flashing lock symbol and beep
several times.
The end of infusion acoustic signal can be set also in the StoP state.
SETTING OF END OF INFUSION ACOUSTIC SIGNAL
NOTES
1. In the OFF state, by pressing the button: the pump
starts the mode for selecting the end of infusion acoustic
signal.
2. When the value flashes, select a new value using the
and buttons.
Selecting OFF disables the warning and the end of infusion
acoustic signals; selecting on activates the warning of end
of infusion acoustic signals, which will go off 5 mins. and
10 mins. before the end of the infusion, and the end of
infusion acoustic signal.
3. Do not press any button for 10 seconds, and the setting
phase will end. The flashing displayed value becomes
fixed and then OFF is displayed.
4. Press the
button before the OFF indication is displayed
(while the value of the end of infusion acoustic signal is still
flashing) to pass to the setting of the successive parameter
(this function is active and is displayed only if the pump
is being programmed in flow rate mode; if the pump has
been programmed in time mode you will go straight to
the next step): SETTING THE NUMBER OF INFUSION
SITES (see page 37).
5. Press the
button while the value of the previous setting
is still flashing to pass to the setting of the next parameter:
SETTING THE PARTIAL VOLUME (see page 39).