This function is active only if the pump has been set in flow rate mode.
In case it is required to administer a drug volume by dividing it in multiple
infusion sites, called Infusion Points (IP), the pump allows you to select up to
5 sites and make partial delivering in succession.
To set this parameter, press the button again.
You can select the number of infusion sites IP (while the number of sites is
flashing) in which the drug volume is to be administered; the selection is done
using the
and buttons; select the number of sites from 1 to 5:
- Select 1 to administer the drug contained in the syringe at one single
infusion site;
- Select 2 to administer 50% of the drug contained in the syringe at one site
and then, the other 50% at another infusion site;
- Select 3 to administer the drug contained in the syringe at 3 different infusion
sites (33% at each site) and so on. Select 4 (25% at each site) and 5 (20%
at each site).
SETTING THE NUMBER OF INFUSION SITES
1 - The first fraction of the drug volume contained in the
syringe will be administered at one infusion site. Then,
the pump will stop and the display will show End, and
it will emit, at the same time, an intermittent acoustic
signal indicating the end of partial infusion until the start
of the next fraction.
2 - Then, the display will show End alternating with 1-4
(fraction 1 out of 4 administered).
SECTION 10
By selecting a number of infusion sites from 2 to 5, for example 4, the pump
will run the administration as follows: