User guide

39
MAN 01/EN/01 CRONO S-PID 50 02/13
SECTION 10
SETTING THE PARTIAL VOLUME
Proceed as follows:
1. The display shows a flashing value for the volume,
preceded by cc, which indicates the unit of volume
(1 cc = 1 ml)
2. Press the button to decrease the value, and the
button to increase it. Each change is indicated by a beep.
3. Do not press any button for 10 seconds and the setting
phase will end. The display will show P,cc.
4. The pusher is automatically positioned at the configured
partial volume value. An intermittent beep is emitted while
it does so, and the pump displays -- in real time -- the actual
volume corresponding to the pusher position.
5. When the pusher is in the correct position the display
changes to OFF.
The partial volume function is used when the therapy requires an infusion with
less than 50 ml.
The partial volume can be set from 1 cc to 50 cc in increments of 1 cc.
To set this parameter, press the
button again while the previous setting
is flashing.
The partial volume function can only be set before the start of a new infusion,
either partial or complete (50 ml).
PROG
PROG