User guide
40
MAN 01/EN/01 CRONO S-PID 50 02/13
SECTION 10
NOTES
WARNINGS
• This operation must not be carried out with the infusion set connected to
the patient.
• A partial volume cannot be set while an infusion is in progress.
• The partial volume setting remains in the pump’s memory even if the
battery is removed.
• If the battery is removed when the pump is set to OFF/StoP, the partial
volume remains in the memory and the pusher is not withdrawn.
• If the battery is removed when the pump is set to ON, the pusher returns
to the infusion start position for recalibration, and then repositions itself at
the stored partial volume.
• The partial volume setting is automatically stored in the pump’s memory.
• At the end of the infusion, the pusher returns to the position corresponding
to the partial volume setting.
• The partial volume setting can be interrupted by pressing the
and
buttons simultaneously.
- If the pusher is still advancing, the pump switches off (the display shows StoP)
and the pusher remains where it was when the infusion was interrupted: the
partial volume setting is not stored and the previous value remains in memory
- If, however, the pusher was in the process of being withdrawn, the display
alternates between OFF and P,cc. The only possible operation is to continue
the withdrawal of the pusher, by pressing the
button. The pusher withdraws
to the position of the partial volume setting.
• Press the
and buttons simultaneously while P,cc it is shown to
cancel the storing of the partial volume.
When the device is set to ON the following parameters can be set:
1 - Delivery time (if time mode has been selected upon battery insertion).
2 - Flow rate (if flow rate mode has been selected upon battery insertion).
MAKING SETTINGS IN THE ON STATE