User guide
41
MAN 01/EN/01 CRONO S-PID 50 02/13
SECTION 10
SETTING THE DELIVERY TIME
This function is available only if the “t” (time) function has
been selected upon battery insertion.
Time value can be set from 30’ to 500 h
- From 30 minutes to 1 hours in increments of 5 minutes
- From 1 hours to 50 hours in increments of 15 minutes
- From 50 hours to 500 hours in increments of 120 minutes
(2 hours).
You can change the time parameter during an infusion.
Procedure:
1 - Switch on the pump by pressing the
button.
2 - Press the
button and the pump allows setting the
infusion delivery – the time display begins to flash.
3 - You can change the parameter using the
or
buttons while the value is flashing.
4 - Not pressing these buttons for about 9 seconds or
pressing the
button will mean that the device will
memorise the parameter that has been selected and
the infusion recommences.
NOTES
• If the settings lock is active the delivery time setting is not accessible; if any
attempts are made to change the parameter then the display will show the
flashing lock icon and the device will give an intermittent acoustic signal.
• If you keep pressing either the or the buttons, it is possible to change
delivery time quickly.
If a partial volume is used, the pump can run an infusion in less time compared
to the minimum time specified (30 minutes).
WARNING
PROG