User guide

42
MAN 01/EN/01 CRONO S-PID 50 02/13
This function is available only if the “F” (flow rate) function
has been selected upon battery insertion.
Flow rate value can be set from 0.1 ml/h to 100 ml/h
- From 0.1 ml/h to 1 ml/h in increments of 0.01 ml/h;
- From 1 ml/h to 10 ml/h in increments of 0.1 ml/h;
- From 10 ml/h to 100 ml/h in increments of 1 ml/h.
You can change the flow rate parameter during an infusion.
Procedure:
1 - Switch on the pump by pressing the
button.
2 - The display shows the selected number of infusion sites
and then the delivery time or the flow rate.
3 - Press the
button and the pump allows setting the
flow rate: the display shows the flashing value of the
flow rate.
4 - You can change the parameter using the
or
buttons while the value is flashing.
5 - Not pressing these buttons for about 9 seconds or
pressing the
button will mean that the device will
memorise the parameter that has been selected and
the infusion recommences.
SETTING THE FLOW RATE
When the settings lock is on the flow rate setting is not accessible; if any
attempts are made to change the parameter then the display will show the
flashing lock symbol and beep several times.
If you keep pressing either the or the buttons, it is possible to change
the flow rate quickly.
If you press the button, the display toggles between the delivery time
and the flow rate and vice versa.
NOTES
SECTION 10
PROG
ml/h