User guide
44
MAN 01/EN/01 CRONO S-PID 50 02/13
PRIMING THE INFUSION LINE
SECTION 10
The priming function allows filling the infusion line with the drugs contained
in the reservoir.
The volume available for priming is 1.5 ml.
The priming function is enabled when you switch on the device and the pusher
is in the infusion start position or at the start of an infusion fraction (Infusion
Point), regardless of whether the settings lock is on.
The priming procedure is as follows:
1. Turn on the device by pressing the
button.
2. The display shows Pr. There are three options:
a. Postpone the priming.
b. Cancel the priming.
c. Carry out the priming.
a. Postpone the priming
Wait 10 seconds, the pump will turn off automatically.
b. Cancel the priming
Press the
button: the pump begins the infusion and
the display shows the time remaining until the end of the
infusion.
c. Carry out the priming
Press and hold down the
button: the pump delivers the
priming dose until you release the button. The display then
shows a flashing letter P in the secondary digits, followed
by the number of ml delivered. When the button is released,
the display shows Pr. The procedure can be repeated up to
a maximum release of 1.5 ml.
Proceed until the infusion set is completely full and a few
drops of the drugs leak out of it.