MAN 01/EN/01 CRONO S-PID 50 02/13
If you keep the button pressed, the pump delivers the priming
dose, giving an acoustic signal every consecutive delivery of 0.5 ml
(i.e. 0.5 – 1.0 – 1.5 ml).
If, after the priming indication is displayed, the buttons are not pressed again
for 10 seconds, the display shows OFF.
The priming function can be interrupted by releasing the
button. The
display shows Pr again , and you again have the choice of postponing,
cancelling or performing the priming function as described above.
At the start of the administration of each Infusion Point the priming function
is available again.
Do not prime the infusion set with the tube connected to the patient.
The priming function must only be performed with the reservoir attached to
the infusion set before inserting the needle into the infusion site.
Before beginning an infusion, check that there are no air bubbles in the
infusion line, expelling any that are found. Alternatively, use a vented filter.