• If you keep the button pressed, the pump delivers the priming
dose, giving an acoustic signal every consecutive delivery of 0.5 ml
(i.e. 0.5 – 1.0 – 1.5 ml).
• If, after the priming indication is displayed, the buttons are not pressed again
for 10 seconds, the display shows OFF.
• The priming function can be interrupted by releasing the
button. The
display shows Pr again , and you again have the choice of postponing,
cancelling or performing the priming function as described above.
• At the start of the administration of each Infusion Point the priming function
is available again.
• Do not prime the infusion set with the tube connected to the patient.
• The priming function must only be performed with the reservoir attached to
the infusion set before inserting the needle into the infusion site.
• Before beginning an infusion, check that there are no air bubbles in the
infusion line, expelling any that are found. Alternatively, use a vented filter.