User guide

46
MAN 01/EN/01 CRONO S-PID 50 02/13
SECTION 10
END OF INFUSION
Ten minutes before the end of the infusion (only if AL is
active), the device gives an intermittent acoustic signal lasting
2 seconds. This signal is repeated twice at 5 minutes from
the end of the infusion and at the end of it: the display shows
the End message.
After a few seconds, the pusher starts withdrawing until it
reaches the start position of the infusion.
When the withdrawal is complete, the display shows OFF
and the pump is ready for a further infusion.
The withdrawal time for a 50 cc syringe is about 6 minutes and is proportionately
less for lower volumes.