WITHDRAWING THE PUSHER
1. Stopping an infusion before the end
This function allows the interruption of an active infusion,
withdrawing the pusher to the start position of the infusion.
To carry out a withdrawal, proceed as follows:
Turn off the pump by pressing the
and the buttons
simultaneously.
Press the
and the button simultaneously: the
display shows End for 10 seconds and then begins to
withdraw the pusher.
During the 10 seconds that the display shows End you
can cancel the withdrawal request by pressing the
and the
buttons together.
2. Withdrawal of the pusher at the end of the infusion
At the end of the infusion the display shows the End
message and the pump will emit an acoustic signal for
a few seconds.
The pusher remains stationary at the end-infusion position
for around 10 seconds, after which it begins to withdraw
until it reaches the start-infusion position.
When the withdrawal is complete, the display shows OFF
and the pump is ready for a further infusion.
Pusher in motion
While the pusher is in the process of being withdrawn,
the display shows the “pusher continuous withdrawal
indication.