User guide

56
MAN 01/EN/01 CRONO S-PID 50 02/13
WARNING
SECTION 11
Before filling the reservoir
Unscrew and screw back the piston rod to facilitate its unscrewing after you
have filled the reservoir.
Filling the reservoir
The liquid must be aspirated slowly.
Do not fill the reservoir more than the maximum level allowed.
The rod must be unscrewed with a fairly rapid movement.
Insertion of the reservoir into the pump
To avoid any leakage of the pharmaceutical while the reservoir is being inserted
into the pump you can use the infusion set, as an alternative to the Luer- Lock
cap indicated on page 51.
When making the connection, avoid exerting any pressure on the reservoir
walls, because this could cause liquid to leak past the piston rings.
While filling the reservoir and inserting it into the pump, a small leakage might
occur between the first and second rings on the rubber piston. This does
not compromise either the correct working of the reservoir or the delivery of
the drug.