User guide
59
MAN 01/EN/01 CRONO S-PID 50 02/13
The images refer to the Neria
TM
infusion set
from Unomedical, a Convatec Company.
Disinfect the infusion site following the
instructions of the relevant medical personnel.
Ensure that the area of the infusion site is dry
before inserting the subcutaneous needle.
Connect the infusion set to the reservoir.
Hold the infusion set by the wings. Prime
the infusion line manually or use the priming
function of the pump. Ensure there are no air
bubbles in the infusion line.
When you are priming the infusion line and
are preparing to insert the needle below the
skin, hold the set with the needle pointing
downwards to ensure that none of the drug
can come into contact with the protecting
adhesive paper.
Remove the protective adhesive paper.
WARNING
SECTION 11