User guide

82
MAN 01/EN/01 CRONO S-PID 50 02/13
APPENDIX 6
Guide and declaration by the manufacturer - electro-magnetic immunity
CRONO S-PID 50 is designed to operate in the electro-magnetic environment specified
below. The customer or user of the CRONO S-PID 50 must ensure that it is operated in
such an environment.
Immunity
test
IEC 60601 test level
Level of
compliance
Electromagnetic
environment - guide
Radiated
immunity
80-2,500 MHz 10V/m
AM 80% 1 KHz
10V/m
20-80 MHz 10V/m
AM 80% 1 KHz
10V/m
Interference could occur in the
vicinity of devices marked with the
following symbol:
Recommended separation distance between mobile and portable radio
communication devices and the CRONO S-PID 50.
CRONO S-PID 50 is designed to operate in an electro-magnetic environment in
which radiated RF disturbances are under control. The customer or user of the
CRONO S-PID 50 can help prevent electro-magnetic interference by ensuring a minimum
distance between mobile and portable communication devices using RF (transmitters) and
the CRONO S-PID 50 , as recommended below, relative to the maximum output power of the
radio-communication devices.
Maximum specified
output power of
transmitter (W)
Separation distance at the transmitter frequency (m)
150 kHz to 80 MHz 80 MHz to 800 MHz
0.01 1.2 0.12
0.1 3.8 0.38
1 12 1,2
10 38 3.8
100 120 12