User guide

106
MAN 01/EN/03 CRONO PAR 50 06/13
APPENDIX 7
Guide and declaration by the manufacturer - electro-magnetic immunity
CRONO PAR 50 is designed to operate in the electromagnetic environment specied be-
low. The client or user of the CRONO PAR 50 must ensure that it is operated in such an
environment.
Immunity
test
IEC 60601 test level
Level of
compliance
Electromagnetic environment -
guide
Radiated
immunity
80-2500 MHz 10V/m
AM 80% 1 KHz
10V/m
20-80 MHz 10V/m
AM 80% 1 KHz
10V/m
Interference could occur in the vici-
nity of devices marked with the fol-
lowing symbol:
Recommended separation distance between mobile and portable radio
communication devices and the CRONO PAR 50
CRONO PAR 50 is designed to operate in an electromagnetic environment in which radia-
ted RF disturbances are under control. The client or user of the CRONO PAR 50 can help
prevent electromagnetic interference by ensuring a minimum distance between mobile and
portable communication devices using RF (transmitters) and the CRONO PAR 50 as recom-
mended below, relative to the maximum output power of the radio-communication devices.
Maximum specied
output power of
transmitter (W)
Separation distance at the transmitter frequency (m)
150 kHz to 80 MHz 80 MHz to 800 MHz
0.01 1.2 0.12
0.1 3.8 0.38
1 12 1.2
10 38 3.8
100 120 12