APPENDIX 7
Guide and declaration by the manufacturer - electro-magnetic immunity
CRONO PAR 50 is designed to operate in the electromagnetic environment specied be-
low. The client or user of the CRONO PAR 50 must ensure that it is operated in such an
environment.
Immunity
test
IEC 60601 test level
Level of
compliance
Electromagnetic environment -
guide
Radiated
immunity
80-2500 MHz 10V/m
AM 80% 1 KHz
10V/m
20-80 MHz 10V/m
AM 80% 1 KHz
10V/m
Interference could occur in the vici-
nity of devices marked with the fol-
lowing symbol:
Recommended separation distance between mobile and portable radio
communication devices and the CRONO PAR 50
CRONO PAR 50 is designed to operate in an electromagnetic environment in which radia-
ted RF disturbances are under control. The client or user of the CRONO PAR 50 can help
prevent electromagnetic interference by ensuring a minimum distance between mobile and
portable communication devices using RF (transmitters) and the CRONO PAR 50 as recom-
mended below, relative to the maximum output power of the radio-communication devices.
Maximum specied
output power of
transmitter (W)
Separation distance at the transmitter frequency (m)
150 kHz to 80 MHz 80 MHz to 800 MHz
0.01 1.2 0.12
0.1 3.8 0.38
1 12 1.2
10 38 3.8
100 120 12