User guide

111
MAN 01/EN/03 CRONO PAR 50 06/13
APPENDIX 10
ACCESSING RESERVED SECTION
You can access additional reserved functions of the
pump, such as: resetting the number of partial infusions,
displaying the number of total infusions, displaying the
firmware version of the pump, setting the occlusion
pressure threshold, setting the date and time.
Procedure:
1 - with the device OFF, press the
button for about 7
seconds, until the display shows rst the number of bolus
doses administered (nd) during the last infusion, followed
by the number of partial infusions (PC) delivered;
2 - press the button (without releasing the button),
and the number of infusions starts to ash. Release both
buttons and press the
button again. In this phase the
arrow indicating PROG is displayed and you can choose
one of the following:
a) reset the partial infusion counter to zero by pressing
the or button, or,
b) continue programming or displaying by pressing the
button, as follows.
3 - Press the button again and the display will show
the total number of infusions.
4 - Press the
button again and the display will show
the rmware version (release).
PROG