APPENDIX 10
ACCESSING RESERVED SECTION
You can access additional reserved functions of the
pump, such as: resetting the number of partial infusions,
displaying the number of total infusions, displaying the
firmware version of the pump, setting the occlusion
pressure threshold, setting the date and time.
Procedure:
1 - with the device OFF, press the
button for about 7
seconds, until the display shows rst the number of bolus
doses administered (nd) during the last infusion, followed
by the number of partial infusions (PC) delivered;
2 - press the button (without releasing the button),
and the number of infusions starts to ash. Release both
buttons and press the
button again. In this phase the
arrow indicating PROG is displayed and you can choose
one of the following:
a) reset the partial infusion counter to zero by pressing
the or button, or,
b) continue programming or displaying by pressing the
button, as follows.
3 - Press the button again and the display will show
the total number of infusions.
4 - Press the
button again and the display will show
the rmware version (release).
PROG