User guide

14
MAN 01/EN/03 CRONO PAR 50 06/13
SECTION 3
TECHNICAL CHARACTERISTICS
Pump dimensions 3.31 x 2.17 x 1.65 in (84 x 55 x 42 mm).
Weight 4.47 oz. (139 g), including battery.
Battery Lithium CR 123A 3V
(battery life of about 100 infusions).
Single-use reservoir With a 50 ml capacity and a “Luer-Lock”
universal security attachment.
Volumes that can be
administered
Selectable, from 1 to 50 ml in 1 ml
increments.
Flow rates in FrEE mode
(F1, F2, F3)
Programmable from:
• 0.05 ml/h to 1.00 ml/h in 0.01 ml/h
increments
• 1.00 ml/h to 3.00 ml/h in 0.02 ml/h
increments;
• 3.00 ml/h to 5,00 ml/h in 0.05 ml/h
increments;
• F2 and F3 may also be programmed
to 0.00 ml/h or to oFF (ow rate display
disabled).
Selectable ow rates in
FrEE mode
3, all programmable.
Flow rates in Auto mode
(F1, F2, F3)
Programmable from:
• 0.05 ml/h to 1.00 ml/h in 0.01 ml/h
increments;
• 1.00 ml/h to 3.00 ml/h in 0.02 ml/h
increments;
• 3.00 ml/h to 5.00 ml/h in 0.05 ml/h
increments;
Selectable ow rates in
Auto mode
4, of which 3 are programmable and 1
is xed at 0 ml/h.
Bolus dose volume Programmable from 0.00 to 2.00 ml in
0.020 ml increments.
Interval between bolus doses Programmable from:
• 5 minutes to 1 hour in 5 minute incre-
ments;
• 1 hour to 24 hours in 15 minutes incre-
ments.
It is possible to disable this function by
setting no,Lt (see page 50 and 73).