User guide

25
MAN 01/EN/03 CRONO PAR 50 06/13
SECTION 6
SETTINGS LOCK
WARNINGS
The CRONO PAR 50 pump has 2 access congurations:
- L0 (unlocked): in this conguration you can use the control buttons to access
all of the settings and parameters, and control all of the operational functions;
- L1 (locked): in this conguration you can use the control buttons to control
the operational functions, except for the time. When the pump is set to L1 the
display shows the lock indicator ( indicator displayed).
Before attempting to modify any of the settings, ensure that the selected access
level of the pump is L0 ( indicator not displayed).
To lock or unlock the settings, do the following:
1. with the pump set to either OFF or StoP, press the
button for about 6
seconds: the display shows the number of bolus doses administered during the
last infusion, followed by the number of infusions that have been completed;
2. without releasing the
button, also press the button: the display shows
either L0 or L1, with 0 or 1 ashing;
3. you can now select whether to lock or unlock the settings by pressing either
or .
• This settings access level (L0 and L1) remains in memory even if the battery
is removed.
• When the settings access is L1 (locked), any attempt to access the locked
options will cause the pump to beep intermittently and display the "lock"
indicator.