SECTION 6
SETTINGS LOCK
WARNINGS
The CRONO PAR 50 pump has 2 access congurations:
- L0 (unlocked): in this conguration you can use the control buttons to access
all of the settings and parameters, and control all of the operational functions;
- L1 (locked): in this conguration you can use the control buttons to control
the operational functions, except for the time. When the pump is set to L1 the
display shows the lock indicator ( indicator displayed).
Before attempting to modify any of the settings, ensure that the selected access
level of the pump is L0 ( indicator not displayed).
To lock or unlock the settings, do the following:
1. with the pump set to either OFF or StoP, press the
button for about 6
seconds: the display shows the number of bolus doses administered during the
last infusion, followed by the number of infusions that have been completed;
2. without releasing the
button, also press the button: the display shows
either L0 or L1, with 0 or 1 ashing;
3. you can now select whether to lock or unlock the settings by pressing either
or .
This settings access level (L0 and L1) remains in memory even if the battery
is removed.
When the settings access is L1 (locked), any attempt to access the locked
options will cause the pump to beep intermittently and display the "lock"
indicator.