SECTION 7
The displayed error messages (from Er,1 to Er,11 and OCCL) are accompa-
nied by an acoustic signal and the system stops.
Following the display of error message Er,1 in the:
- ON condition (Auto version) after you press the
button, the pump recom-
mences the delivery of the drug, but only using the lowest of the programmed
ow rates;
- OFF condition, after you press the
button, the pump will be set back to
OFF;
- Battery insertion condition, after you press the
button, the pump will be
initialised (as described on page 30).
The display will show the clock icon and the ashing dashes on the two
secondary digits until the time is reset.
It is necessary to reset the hours and minutes by accessing the settings as
described in the “PUMP SETTINGS SEQUENCE” section on pages 38 and 59.
If at the time when Er,1 is signalled the hourly ow value is set to F0, the
pump will NOT administer the drug until it is switched off and on again.
If the time is not reset, every time the pump starts or resumes an infusion,
it will signal Er,1 and administer the drug for the whole delivery time at the
minimum ow among the 3 set values.
In the event that it is not possible to reset the time from the settings menu of
the pump, contact the CANÈ S.p.A. Technical Support Service.
Error messages Er,2 and Er,7 are accompanied by the ashing red LED.
Following the display of error message Er,8 and the successive initialisation,
the system re-imposes the factory settings in FrEE mode (see pages 32 and
53): in this event the pump settings prescribed by the doctor should be re-
entered.
The settings prescribed by the doctor must be noted on a copy of the patient
settings record sheet on page 110 or the notes sheet on page 113.