User guide

31
MAN 01/EN/03 CRONO PAR 50 06/13
To be able to select the mode of the pump, it must:
be set to OFF (at the beginning of a full or partial infusion);
have the settings lock set to L0, ("lock" indicator off).
SECTION 8
Procedure:
1 - press the button for about 10 seconds, until the
display subsequently shows the number of bolus doses
administered during the last infusion, the number of com-
pleted infusions and nally the SEt message;
2 - when the message SEt is displayed, release the
button and press the
. The display shows Auto or
FrEE ashing message and the arrow points to PROG
demonstrating that the pump is ready to be programmed;
3 - press the
button to switch from Auto to FrEE, or
press the
button to switch from FrEE to Auto;
4 - if you have set FrEE go to the next page. If you have
set Auto go to page 53.
WARNINGS
Always verify that the clock is correctly set when switching pump mode
from FrEE to Auto.
When the device is set to StoP it is not possible to change the pump mode,
and the pump displays Err.
PROG
PROG