To be able to select the mode of the pump, it must:
• be set to OFF (at the beginning of a full or partial infusion);
• have the settings lock set to L0, ("lock" indicator off).
SECTION 8
Procedure:
1 - press the button for about 10 seconds, until the
display subsequently shows the number of bolus doses
administered during the last infusion, the number of com-
pleted infusions and nally the SEt message;
2 - when the message SEt is displayed, release the
button and press the
. The display shows Auto or
FrEE ashing message and the arrow points to PROG
demonstrating that the pump is ready to be programmed;
3 - press the
button to switch from Auto to FrEE, or
press the
button to switch from FrEE to Auto;
4 - if you have set FrEE go to the next page. If you have
set Auto go to page 53.
WARNINGS
• Always verify that the clock is correctly set when switching pump mode
from FrEE to Auto.
• When the device is set to StoP it is not possible to change the pump mode,
and the pump displays Err.