User guide

38
MAN 01/EN/03 CRONO PAR 50 06/13
FrEE
Proceed as follows:
1. with the pump set to OFF or StoP press the
button
for a few seconds: the display shows the ashing BOLUS
DOSE VOLUME indication;
2. to change the value, within 20 seconds press the
button to decrease it, or the
button to increase it. Each
change is indicated by a beep;
3. if you do not press any button for 20 seconds, the settings
session is terminated and the message OFF or StoP is
displayed;
4. press the
button before the OFF or StoP is displayed
(while the bolus dose volume is still ashing) to continue
to the setting of the next parameter: SETTING THE
INTERVAL BETWEEN BOLUS DOSES.
PUMP SETTINGS SEQUENCE WITH THE PUMP SET TO OFF OR StoP
To change the settings the pump must:
be set to OFF or StoP;
have the settings lock off (i.e. set to L0).
The bolus dose volume can be set from 0.00 to 2.00 ml in increments of 0.022 ml.
PROG
If the bolus dose volume is set to 0.00 ml, pressing the button during the
infusion will display the Er,d error message.
NOTES
WARNING
If you keep either the or button pressed, it is possible to quickly
change the bolus dose volume.
The bolus dose setting is automatically stored in the pump’s memory.
If you press the button when the display indicates d0,02, the value chan-
ges to d0,00: this setting disables the bolus dose function.
If the settings lock is set to L1 (the display is showing the "lock" indicator),
when you press the button, the pump displays the current settings, which
cannot be changed.
SETTING THE BOLUS DOSE VOLUME
PROG