User guide

41
MAN 01/EN/03 CRONO PAR 50 06/13
FrEE
FrEE
SETTING THE SECOND AND THIRD FLOW RATES (F2 AND F3)
The ow rate can be set from 0.00 to 5.00 ml/h as follows:
• from 0.00 ml/h to 1.00 ml/h in increments of 0.01 ml/h;
• from 1.00 ml/h to 3.00 ml/h in increments of 0.02 ml/h;
• from 3.00 ml/h to 5.00 ml/h in increments of 0.05 ml/h.
The second and third ow rates are set in the same manner
as the rst.
After you have set the third ow, press the
button before
the OFF or StoP indication is displayed (while the F3 hourly
ows value is still ashing) to continue to the setting of the
next parameter: PARTIAL VOLUME.
• If you keep either the
or button pressed, it is possible
to change the ow rate quickly.
• If the ow rate is programmed to 0.00 ml/h, the device
will not administer the drug via ow, but only through the
administration of bolus doses.
• it is also possible to set the oFF condition for ows F2 and/
or F3, so that the ows will not be displayed, and therefore
will not be selectable, when the pump is ON.