User guide

43
MAN 01/EN/03 CRONO PAR 50 06/13
FrEE
FrEE
NOTES
WARNINGS
• This operation must not be carried out with the infusion set connected to
the patient.
• A partial volume cannot be set while an infusion is in progress.
• The partial volume setting remains in the pump’s memory even if the bat-
tery is removed.
• If the battery is removed when the pump is set to OFF, the partial volume
remains in memory and the pusher is not withdrawn.
• If the battery is removed when the pump is set to ON, the pusher returns
to the infusion start position for recalibration, and then repositions itself at
the stored partial volume.
• The partial volume setting is automatically stored in the pump’s memory.
• At the end of the infusion, the pusher returns to the position corresponding
to the partial volume setting.
• The partial volume setting can be interrupted by pressing the
button: the
pump switches off (the display shows OFF or StoP) and the pusher, if it was
moving forward, remains where it was when the interruption occurred. The
partial volume setting is not stored and the previous value remains in memory.
If, however, the pusher was in the process of being withdrawn, the display
shows OFF and P,cc alternately. The only possible operation is to continue
the withdrawal of the pusher, by pressing the
button. The pusher withdraws
to the position of the partial volume settings.
• When the device is set to StoP it is not possible to change the partial volume,
and the pump displays Err.
SECTION 11