User guide

48
MAN 01/EN/03 CRONO PAR 50 06/13
FrEE
SELECTING THE FLOW RATE
SECTION 11
This function allows the patients to change the ow during the day based on
their demands, choosing among the 3 available ow rates pre-programmed
by the doctor.
Procedure:
1 - with the pump on, press the
button;
2 - the display shows the currently set ow rate ashing with
the arrow indicating the programming mode;
3 - press the
button again, and the display shows the
other available ow values;
4 - after selecting the desired ow, do not press the
button again, and after a few seconds the value is saved in
the memory and the pump will proceed with the infusion at
the newly selected ow rate;
5 - during this selection the
and buttons are disabled.
NOTES
• If F2 and/or F3 are set to 0.00, the pump allows the selection of these ow
values as an alternative to F1, with no administering being performed as the
ow is equal to 0; selecting a 0 ow can be convenient to intentionally interrupt
the administration for a set period, for instance during the night.
• If F2 and F3 are set to oFF, the pump allows the display and use of the F1
ow only.
The preset ow selection function is available even when the settings lock is
on; to disable this option, it is necessary to set the F2 and F3 ows
to oFF in
order to disable them, forcing the pump to use a unique ow rate.
WARNING
PROG
ml/h
PROG
ml/h