User guide

61
MAN 01/EN/03 CRONO PAR 50 06/13
Auto
SETTING THE TIME
Proceed as follows:
1. with the pump set to either the OFF or StoP conditions,
press the
button for about 4 seconds: the display shows
the time;
Pressing the
button enables the setting of the time: the
hour value begins to ash.
2. press the
button to increase the value; press the
button to decrease it. Each change in value is indicated by
a brief beep;
3. do not press any button for 20 seconds, and the setting
phase will end. The ashing displayed value becomes xed
and then the OFF or StoP indication is displayed;
4. press the
button before the OFF or StoP is displayed
(while the hour value is still ashing) to continue to the setting
of the minutes;
5. press the
button to increase the minutes; press the
button to decrease them. Each change in value is indicated
by a brief beep;
6. do not press any button for 20 seconds, and the setting
phase will end. The ashing displayed value becomes xed
and then the OFF or StoP indication is displayed.
7. press the
button before the OFF or StoP indication is
displayed (while the minutes are still ashing) to continue to
the setting of the next parameter: SETTING THE BOLUS
DOSE VOLUME.
SECTION 14
NOTE
If you keep pressing either the or the button, it is possible to quickly
change the hours and minutes.
PROG
PROG
PROG