SETTING THE FIRST FLOW RATE (F1)
Proceed as follows:
1. the display shows the ashing value of the rst ow rate;
2. press the
button to increase the value; press the
button to decrease it. Each change is indicated by a beep;
3. if you do not press any button for 20 seconds, the settings
session is terminated and the message OFF or StoP is
displayed;
4. press the
button before the OFF or StoP indication
is displayed (while the F1 ow rate value is still ashing) to
continue to the setting of the next parameter: SETTING THE
SECOND FLOW RATE (F2).
If you keep pressing either the or buttons, it is possible to change
the ow rate quickly.
Every ow rate setting is automatically stored in the pump’s memory.
NOTES
SECTION 14
The ow rate can be set from 0.05 to 5.00 ml/h as follows:
from 0.05 ml/h to 1.00 ml/h in increments of 0.01 ml/h;
from 1.00 ml/h to 3.00 ml/h in increments of 0.02 ml/h;
from 3.00 ml/h to 5.00 ml/h in increments of 0.05 ml/h.