Auto
ASSIGNING THE FLOWS OVER THE DAILY 24-HOUR PERIOD
Proceed as follows:
1. the rst two digits of the display show the hour of the day
(24-hour clock) while the last two show the selected ow
for the corresponding hour of the day. The rst three digits
are steady, but the last numeric digit, after the F (0, 1, 2 or
3) indicating the ow, is ashing;
2. press the
button to select the hour of the day;
then press the
button to select the required flow
corresponding to the selected hour of the day. Each change
is indicated by a beep. Press the
button again to change
the display to the next hour: press the
button to select
the required ow. Proceed in the same way to set the ows
for each successive hour;
3. do not press any button for 20 seconds, and the setting
phase will end. The ashing displayed value becomes xed
and then the OFF or StoP indication is displayed;
4. press the
button before the OFF or StoP indication
is displayed (while the hourly ow value is still ashing) to
continue to the setting of the next parameter: SETTING
THE PARTIAL VOLUME.
SECTION 14
NOTE
PROG
PROG
You can set a different ow rate for every hour of the day, choosing between
the four available rates (F0, F1, F2, F3).
If you set the F0 ow rate for any hour, during that hour no drug will be
administered, as the ow rate will be 0.00 ml/h: the display shows the remaining
time of the infusion, and pressing the
button will temporarily display F0
with the value of 0.00 ml/h.