User guide

67
MAN 01/EN/03 CRONO PAR 50 06/13
Auto
SETTING THE PARTIAL VOLUME
Proceed as follows:
1. the display shows a flashing value for the volume,
preceded by cc , which indicates the unit of volume (1 cc
= 1 ml);
2. press the
button to decrease the value, and the
button to increase it. Each change is indicated by a beep;
3. do not press any button for 20 seconds and the setting
phase will end; the display will show P,cc;
4. the pump moves the pusher to the position which
corresponds to the set volume: an intermittent acoustic
signal is emitted while it does so, and the pump displays - in
real time - the actual volume corresponding to the pusher
position;
5. when the pusher is in the correct position the display
shows OFF.
SECTION 14
PROG
PROG
The partial volume function is used when the treatment requires an infusion
with less than 50 ml.
The partial volume can be set from 1 cc to 50 cc in increments of 1 cc. To set
this parameter, press the
button again.
The partial volume function can only be set before starting a new infusion,
either a complete one (50 ml) or a partial one.