User guide

71
MAN 01/EN/03 CRONO PAR 50 06/13
Auto
SECTION 14
Every time the pump administers 0.5 ml of drug, it emits an audible signal
and pauses for about one second.
Continue with the priming until the drug arrives at the end of the infusion set.
There are no time limits to the priming operation.
If, after the priming indication is displayed, the buttons are not pressed
again for 10 seconds, the display shows OFF.
Do not prime the infusion set with the tube connected to the patient.
The priming function must only be performed with the reservoir attached to
the infusion set and before inserting the needle into the infusion site.
Before beginning an infusion, check that there are no air bubbles in the infu-
sion line, expelling any that are found. Alternatively, use a vented lter.
While the displays ashes Pr it is not possible to select the programming
mode (FrEE/Auto) or change the partial volume.
NOTES
WARNINGS
c. Perform the priming
Press and hold down the
button to deliver the priming
dose; the priming function can be interrupted by releasing
the button. The display shows a ashing letter Pr followed
by the number of millilitres delivered. The display then
shows Pr again. In this way it is again possible to postpone,
cancel or perform the priming function. The procedure can
be repeated until a maximum of 1.5 ml has been delivered
in total.
At the end of this phase, if the infusion is not started by
pressing the
button, the pump returns to OFF.