User guide

72
MAN 01/EN/03 CRONO PAR 50 06/13
Auto
PUMP SWITCHED ON
During the infusion the time indicated on the display is shown
in hours (with a ashing letter h), or in hours and minutes
with the ashing drop icon. If the time "to go" is less than
one hour, the time is shown in just minutes, and the ashing
“Minute” indicator is displayed.
If you press the
button, the display temporarily toggles
between the "time to go" and the programmed ow rate.
If you have selected the F0 ow, the display shows F0 and
0,00. In this event the drug can only be administered by
means of a bolus dose request.
1. Press the
button; the pump gives a beep and
commences to administer the bolus dose, during which a
ashing letter d is shown on the secondary display, while
the volume of the bolus dose being administered is shown
on the main display. During the bolus dose administration,
the pump beeps and pauses for about 1 seconds for every
0.11 ml of drug administered.
2. At the end of the bolus dose administration, the pump
beeps and the display shows the time remaining of the
original infusion, or, if the setting is for a ow rate of 0.00
ml/h, it shows F0,00.
BOLUS DOSE
SECTION 14
ml/h
The bolus is a supplementary dose of the drug that the patient can request
when it is needed. If the bolus dose volume has been set to 0.00 ml, the bolus
dose cannot be administered.
Administration of bolus dose
Bolus doses can only be administered while the pump is in ON condition.
The procedure for their administration is as follows:
ml/h