User guide

73
MAN 01/EN/03 CRONO PAR 50 06/13
Auto
The bolus dose delivery can be interrupted at any time by pressing the
button.
If the bolus dose volume is set to 0.00 ml, pressing the
button will display the Er,d error message.
The administration of bolus doses is limited by the interval
between doses (temporal limitation).
By setting no,Lt you can disable this limitation.
Bolus doses cannot be administered during the interval
period observed by the function.
During the limitation period, if you request a bolus dose,
the display shows Er,t (temporal limitation between active
bolus doses).
SECTION 14
NOTE
NOTE
ml/h
RESTRICTIONS ON THE ADMINISTRATION OF BOLUS DOSES
DISPLAYING THE NUMBER OF ADMINISTERED BOLUS DOSES
DURING AN INFUSION AND THE NUMBER OF INFUSIONS
1 - To check the number of bolus doses administered by the
patient during an infusion while the pump is OFF or StoP,
press and hold the
button for a few seconds and the
value will be displayed (nd = number of doses).
2 - To check the number of infusions delivered by the device
(PC = partial counter) while it is OFF or StoP, press and
hold the
button for about 8 seconds.
If all the hourly ows are set to F0, the display shows F0,00
with a ashing F. Pressing the
button temporarily displays
F0 with 0.00 ml/h.
ml/h