User guide

75
MAN 01/EN/03 CRONO PAR 50 06/13
Auto
SECTION 14
NOTES
• While the display shows End, the withdrawal function can
be cancelled by pressing the
button. The display shows
StoP.
• The withdrawal function can be interrupted by pressing
the
button. The display then alternates between End
and OFF. At this point the only active button is the
button. When you press again the pump recommences the
withdrawal of the pusher.
Do not remove the reservoir until the pusher has been withdrawn to the infusion
start position.
WARNING
SWITCHING OFF THE PUMP
To switch off the device, press the button. The display
will show StoP. If the pump is switched off during an infu-
sion, the device will emit a series of 10 short beeps every
10 seconds, and the display will ash the StoP message.
To suppress the audible signals, press the
button.
The pusher withdrawal time for a 50 ml volume is approx 6 minutes, and is
proportionately less for lower volumes.
Pusher in motion
While the pusher is in a continuous withdrawal motion, the
display shows the “pusher continuous withdrawal” indi-
cation.
NOTE