User guide

14
TECHNICAL FEATURES
MAN 01/06/USA/p CRONO FIVE 11.10
Pump size 77 x 47 x 29 mm.
Weight 115 g (battery included).
Battery CR 123A 3V Lithium battery (battery
life: about 6 months).
Syringe Dedicated 10 and 20 ml capacity with
luer-lock connector.
Administration volumes From 1 ml to 20 ml with increments of
1 ml.
Priming Available only at the beginning of an
infusion or partial infusion. The maxi-
mum deliverable volume is 1.5 ml with
0.5 ml increments.
Flow rate Programmable from 5 to 5000 mcl/h
with the following increments:
- from 5 to 1000 mcl/h with 5 mcl/h
increments;
- from 1000 to 5000 mcl/h with 10
mcl/h increments;
The flow rate can be programmed to 0 ml.
Bolus dose Programmable from 0.01 to 2.00 ml
with 0.01 ml increments:
The bolus dose can be programmed to 0 ml.
Clinician bolus Programmable from 0.01 to 2.00 ml
with 0.01 ml increments:
This function is always locked.
Minimum time between bolus doses Programmable from 5’ to 24 h as
follows:
- from 5' to 1h with increments of 5';
- from 1 h to 24 h with increments of 15';
- this function can be deactivated by
programming 0.
SECTION
2
INFUSION SYSTEM
The pump administers shots of 5 microliters (µl or mcl) for any given flow
rate. The time interval between shots decrease proportionally to increases
in programmed flow rates.