User guide
15
MAN 01/06/USA/p CRONO FIVE 11.10
Number of boluses doses per - from 1 to 12 boluses.
hour - this function can be deactivated by 
programming 0.
Flow accuracy +/-2% (observation period 40 minutes).
Max. occlusion pressure 2.2 bar +/-0.8.
Time to an occlusion alarm Please, see ANNEX 2.
Post-occlusion bolus About 0.4 ml (details are supplied on
ANNEX 3).
Electric circuit  Electric circuit managed by a micro
controller equipped with a dedicated 
software.
Data storage The selected data are automatically 
stored in the pump memory and they 
are not lost when the battery is removed.
Display LCD.
Motor Direct current coreless motor.The 
microcontroller manages the rotation 
through infrared encoder.
Safety circuits To monitor the correct working of the 
device intervening in case of anoma-
lies with acoustic warnings and error 
messages.
Anti free-flow system To avoid a non controlled flow inside 
the delivery set due to the force of 
gravity.
Operating conditions +10°C / +45 °C 
30% / 75% RH
700 hPa / 1060 hPa
Storage conditions -10°C / +60 °C 
10% / 100% RH 
500 hPa / 1060 hPa
SECTION 