User guide

15
MAN 01/06/USA/p CRONO FIVE 11.10
Number of boluses doses per - from 1 to 12 boluses.
hour - this function can be deactivated by
programming 0.
Flow accuracy +/-2% (observation period 40 minutes).
Max. occlusion pressure 2.2 bar +/-0.8.
Time to an occlusion alarm Please, see ANNEX 2.
Post-occlusion bolus About 0.4 ml (details are supplied on
ANNEX 3).
Electric circuit Electric circuit managed by a micro
controller equipped with a dedicated
software.
Data storage The selected data are automatically
stored in the pump memory and they
are not lost when the battery is removed.
Display LCD.
Motor Direct current coreless motor.The
microcontroller manages the rotation
through infrared encoder.
Safety circuits To monitor the correct working of the
device intervening in case of anoma-
lies with acoustic warnings and error
messages.
Anti free-flow system To avoid a non controlled flow inside
the delivery set due to the force of
gravity.
Operating conditions +10°C / +45 °C
30% / 75% RH
700 hPa / 1060 hPa
Storage conditions -10°C / +60 °C
10% / 100% RH
500 hPa / 1060 hPa
SECTION
2