WARNING
This safety system will only work if the
connection between the syringe rub -
ber piston and the pump pusher is
made correctly, as shown in the dra -
wing below.
SECTION
5
31
MAN 01/06/USA/p CRONO FIVE 11.10
SAFETY SYSTEM TO AVOID FREE-FLOW
"Free flow" means an uncontrolled
release of the infusion solution
from the syringe caused by gra -
vity. In order to prevent this from
happening, the tip of the pump
pusher has been knurled, which
part must be inserted inside the
rubber piston of the Crono 10-20
ml syringe, allowing a secure con-
nection to be made with the syrin -
ge piston.
CAUTIONS
Before the drug administration is
started the rubber piston has to be
connected to the pump pusher ,if
this is not the case the following
unwanted effects may occur:
The infusion starts late and is
not performed with accuracy.
The anti free flow function is not
operational.
WARNING
The improper connection between
the syringe piston and the pump
pusher can cause a free flow of
medication, in other words, medi -
cation administered by gravity and
not controlled by the pump, (for
example when the pump is posi -
tioned much higher than the infu -
sion site). Such situation can
cause serious injury or death of
the patient.
NO
YES