DEVICE ACTIVATION
MAN 01/06/USA/p CRONO FIVE 11.10
When you insert the battery the pump starts a
sequence of activation during which:
1.The pump will carry out a self-test with brief
acoustic signals and all symbols will appear
on the screen.
2. Select the syringe size options.
3.The pump’s mechanical pusher will place
itself in the correct starting position and at
the end of this self-adjusting, OFF will
appear on the screen.
NOTE
• The pump is supplied with a battery already
inserted.
• Refer to the relevant paragraph for instruc-
tions how to insert a new battery (see page
66).
• Take out the battery of the pump if it is not
going to be used for a long period of time
(1-2 months).
SELECTING THE SYRINGE SIZE
While syringe size symbol is flashing (the
number “10” or “20”), the selection of syringe
size may be made.
Use the or buttons to make your selec-
tion.
The information regarding the syringe size will
always be shown on the display.
10
20
10
20
10
20
10
20
WARNING
Selecting the syringe size
option can only be carried
out during the device acti -
vation phase, immediatly
after the insertion of the
battery.